GilAir Plus – personal air sampling pump
GilAir Plus is a personal air sampling pump from Sensidyne featuring QuadMode™ technology, covering both low- and high-flow applications within 20–5,100 cc/min. The pump provides constant flow or constant pressure operation without external adapters and is intended for industrial hygiene, environmental monitoring and indoor air quality sampling.
Features and benefits
- QuadModeTM Air Sampling Technology: It’s like 2 pumps in 1 – Performs both high-flow constant pressure and constant flow (450–5100 cc/min) and low-flow constant pressure and constant flow (20–499 cc/min) with a single pump and no external adapters.
- Selectable Automatic Fault Recovery: Ensures sample integrity by attempting to restart the pump in the event of a fault.
- Standard Temperature and Pressure Correction (STP): Provides accurate flow rate and volume measurements under varying environmental conditions (STP model).
- Superior Design and Comfort: Compact and lightweight design (one-third smaller than traditional pumps, weighing less than 59g) maximizes working comfort during extended sampling periods.
- Real-time clock with data logging (up to 16 events): Collects and stores critical sampling data for easy retrieval and analysis.
Specifications
- Flow Range: Constant Flow 20-5100 cc/min without external adapters, Constant Pressure 1-5100 cc/min without external adapters
- Flow Modes: Constant Flow or High and Low Flow, Constant Pressure
- Accuracy: Flow Display – 5% or 3 cc/min of Set Flow, whichever is greater, Constant Flow Control – 5% or 3 cc/min of Set Flow, whichever is greater
Constant Pressure Control – 10% of Back Pressure - Back Pressure Capacity: (8 hours of operation)
5000 cc/min. up to 12″ H2O Back Pressure
4000 cc/min. up to 20″ H2O Back Pressure
3000 cc/min. up to 30″ H2O Back Pressure
2000 cc/min. up to 30″ H2O backpressure
1000 cc/min. up to 35″ H2O backpressure
450–1000 cc/min. up to 40″ H2O backpressure
20–449 cc/min. up to 40″ H2O backpressure - Runtime: Minimum 8 hours
- Flow Error: If flow changes exceed 5% within backpressure specifications, an error message will be displayed. If the error exceeds 30 seconds, the pump will shut down. Selectable automatic error reset allows the pump to attempt to restart every 3 minutes for up to 10 times or hold until manual action is taken.
- Temperature range: Operating 0°C to 45°C, Storage -20°C to 45°C, Charging 5°C to 40°C
- Humidity range: Operating 5–95% RH, non-condensing, Storage 5–98% RH, non-condensing
- Display: Flow rate, sample time, and sample volume under actual conditions
- Controls: Indicators Green and red LEDs
- Dimensions: 10.9 x 6.1 x 6.1 centimeters
- Weight: 581 grams
- Warranty: 2 years from original shipment, 5 years for pump keypad, and 1 year for rechargeable NiMH battery pack
- Power options
Removable rechargeable nickel-metal hydride (NiMH) battery pack, optional alkaline battery pack, or DC power. - Battery level indicator: On-screen icon indicates battery level.
- Docking features: Battery charging, PC interface and automatic calibration.
- Charging time: Less than 3.5 hours.
- ATEX/IEC: Ex – II 1 G; Ex ia IIC T4